How to maintain and maintain the washable test chamber
One:
1 The internal and external structure of the water resistant test chamber is sprayed with a high-grade stainless steel plate. The water inlet is also softened, but the nozzle is not often cleaned so that water stains remain on the nozzle, which may block the nozzle for a long time. Whenever the test is completed, the air of the air spray gun nozzle is used.
2. The passive bearing housing (IPX1X2) is located above the equipment in four groups, and the bearings should be lubricated with lubricating oil every season.
3. Passive (IPX3X4) device on the right side, apply lubricant lubrication every season
4. The passive chain is located at the rear of the equipment. The upper cover must be opened. The chain of the drive chain should be lubricated with the chain oil and the top of the equipment. The season should be lubricated with the chain lubricant.
Note: 1. Chain lubricants should not be lubricated with other butter or other solid grease to prevent the speed sensor from malfunctioning.
2. 60 teeth of the speed sensor, the toothed disc should not be lubricated with any grease.
5. There are 2 sets of water filters on the inlet side of the shaft seat. It should be soaked in NaCl (salt) for 24 hours every year, then rinsed with water.
Two: The steps are as follows:
1. Open the water inlet valve in the closed state.
2. Remove the left and right sides of the transparent hose.
3. Unscrew the black upper cover in the counterclockwise direction, and pull it up to 10CM, then put 2 kg/4 bottles of sea salt into the bottle, then screw the black cover in a clockwise direction. Lock the transparent hose back (the water outlet should not be locked first, then the waste water drain pipe should be connected), and the water and water should be locked back in the exact direction.
4. Stand still for 24 hours, open the water inlet valve, and discharge the waste water pipe to the water hole when the water outlet is terminated and the temporary drain pipe is used.
5. Rinse for 30 minutes.
After 6.30 minutes, complete the water tank into the water pipe and the filter tube.
7 The small filter element on the left side of the water resistant test box equipment depends on the water quality; generally it can be used for six months. If the filter element is visually obscured by the naked eye, it needs to be replaced. When the filter element is replaced, the filter element wrench is turned from the bottom to the top of the filter cup and then unscrewed counterclockwise. When the filter bowl is loosened, do not use a wrench, but hold the filter bowl with both hands until the filter bowl is completely detached, take out the filter bowl, take out the old filter element, pour out the water, and clear the filter bowl liner. , then install the new filter.
(The round hole in the middle of the filter element must be in the center of the filter bowl) in the filter cup, then hold it with both hands, tighten it clockwise, then tighten with a wrench.
8. When cleaning the floor in the tank, use a general floor cleaner. If necessary, brush with a brush.
9. The outer layer of the equipment should be cleaned with water on a normal cotton cloth and wiped with water wax if necessary.
Note: It is strictly forbidden to use liquid wipes such as benzyl alcohol.
10. Control the glass of the compartment to wipe gently with normal cotton fabric or old newspaper.
11. Operate the faceplate with a light cotton cloth (do not touch water).